60
Set the start distance of transform.
[Stop Time]
Set the stop distance of transform.
【
Mode
】
Set the transform mode to bandpass, lowpass@step or lowpass@impluse.
[Set Freq Lowpass]
When the transform mode is lowpass, there are special requirements for frequency step. Press this
button to automatically adjust the start frequency and stop frequency, so as to meet the requirements.
[Window]
Set the window type.
[Max]
Set the maximum value of the window.
[Normal]
Set the normal value of the window.
[Min]
Set the minimum value of the window.
[Imp Width 0.036ns]
Set the impulse width of the window.
[Kaise β]
Set Kaise β of the window within the range of 0-13.
[[Velocity]
Set the transform velocity within the range of 0-1.
[Unit m ft]
Set the distance unit.
Section 10 Technical Specifications
After S5105D/E/F microwave analyzer is stored for two hours at the ambient temperature and warmed
up for 15 minutes after starting, the following performance characteristics should be satisfied in
vector network analyzer mode.
a)
Frequency range:
30kHz - 18GHz (S5105D)
30kHz - 26.5GHz (S51057E)
50MHz - 40GHz (S5105F);
b)
Power level: high, low or manually set;
c)
Power setting step: 1dB;
d)
Port output power (high power) range: -11dBm to +1dBm
e)
Power level accuracy (output -15dBm):
±2.5 dB
10MHz - 26.5GHz
±3.0dB
26.5GHz - 40GHz
f)
Effective direction:
S5105D/E:
≥40 dB
2MHz - 500MHz
≥36 dB
500MHz - 9GHz