Measurement Setting
This chapter will introduce how to make the measurement settings. Normally, following setting operations should be done during a
test. This chapter will introduce these operations in sequence.
Resetting Analyzer
Selecting Measure parameter
Setting frequency range
Setting signal power level
Sweep Setup
Selecting triggering mode
Setting data format and scale
Observing multi-Trace and enabling multichannel
Setting Analyzer display
Configure 4-port Measure
4.1.
Resetting Analyzer
4.1.1.
Reset state
When
[Preset]
key is pressed, the Analyzer will be returned to a default state which is known as reset state, and the reset state for
S3601 Series Vector Network Analyzer is set as follows:
Measure parameter: S11
Frequency setting:
Start frequency: 100kHz
Stop frequency: 3GHz/8.5GHz
CW Time: 2GHz
Power setting:
Measuring port power: 0dBm
Coupled port power: on Attenuation :
automatic attenuation Attenuation
value : 0dB
Sweep Setup:
Sweep type: linear frequency
Sweep time: automatic
Sweep spot number: 201
Segment Sweep Setup:
Number of started segment: 1
Start frequency: 100kHz