10.2.4.
Remote Calibration
Remote calibration refers to control the mechanical calibration of instrument by computer. Click the
【
Remote Ctrl
】
tab, and then
select
【
Calibration
】
in the
【
Remote Cal
】
group to pop up the dialog box of “Calibrate”. Operating procedures of remote calibration
are shown as follows:
1) In the “Set Cal Kit” column of the dialog box, click the corresponding key to select model of calibration kit;
2) Connect the “Open Circuit” of calibration kit to test port of instrument, after that, click the
【
OPEN
】
key in the dialog box. After the
open circuit has been tested, the key text will be marked with green underline;
3) Complete the test of short circuit and load by referring to Step 2;
4) After all calibration standards have been tested, the
【
Finish
】
key in the dialog box will become into normal state. And then press this
key to close the dialog box to complete the calibration.
Figure 10-9
【
Remote Cal
】
Group