background image

Waterproof Scale Operation Instruction

11

X=1~9, the display weight is hold while it is stable and keep 1~9seconds.

7.3 Refresh mode

The screen displays

“FLt2. X ”

. It is the refresh mode after hold function while set

“Flt0 =1

”.

X=0, when unload the weight until it less than 20d, holding is canceled and weight is refreshed.
Y=1~9, when weight is on holding status, the holding will be canceled automatic after preset seconds and

refresh and hold a new weight. Time increases by number: [1= 1second, 9 = 9 seconds].

Step 8: Save the Setting

The screen displays

“ – PASS –”

. This step is for saving all above operation. Press

[→0←]

key to save

the setting and exit to weighing mode.

Step 9:

Configuration is done

Plug the breaker JP1 into “OFF”.

XI. ERROR SIGNAL

There will shows error signal when the scale have some problem.

Error 1:

means calibration isn’t workable, weight is too light or division is too high.

Error 2:

means wrong zero, check the load cell if it is damaged.

Error 3:

means displayed value exceed display range after unit change.

Error 4:

means ISN=0

Error 5:

means overload, loaded weight exceed the full range.

Error 6:

means cannot zero.

Error 7:

means cannot tare.

Error 8:

means cannot M+Print.

Error 9:

means cannot unit conversion.

Error 10:

means not connect the battery or battery damage.

XII.WARRANTY

We guarantee one-year of free maintenance since the date of purchase for any

non-manmade faults in normal working conditions. For maintenance, please send the

equipment with the guarantee card to our sales service.

Attention:

we are always improving the machine, there is not pre-notice if there are

something different comparing with the precious ones.

Summary of Contents for KW1000G-15

Page 1: ...Waterproof Scale Operation Instruction 1 WATERPROOF SCALE Operation Instruction Ver1 0...

Page 2: ...PREPARING TO USE THE SCALE 1 III INTRODUCTION 1 IV LCD DISPLAY SYMBOLS 2 V KEYBOARD FUNCTION 3 VI GENERAL PARAMETER 4 VII CONNECTION OF INTERFACE 6 VIII MEANS OF POWER SUPPLY CHARGING 7 IX COUNTING F...

Page 3: ...our leveling feet to set the level of scale platform 3 Avoid operating the scale in direct sunlight or drafts of any kind 4 Take away any weight that might be on the platform before the scale is switc...

Page 4: ...pacity indication is displayed shortly on the display the internal battery needs to be recharged The scale will power off automatically without recharging after the low battery symbol shows upon for 2...

Page 5: ...Capacity Shining is charging Kg lb lb oz g PCS Weighing unit HI OK LO Checking weighing Front panel Keyboard V KEYBOARD FUNCTION KEYS DESCRIPTIONS Turn on off The indicator will be turned on when pre...

Page 6: ...ANDBY MODE Set Pr AB It is parameter of LED display is closed for saving power A for Standby mode B for the effective time A 0 means LED is lighted all the time A 1 means Only STB indication is lighte...

Page 7: ...n is working X 0 means the indicator will keep checking and alert even if the reading is unstable X 1 means the indicator will alert after the reading is stable X 2 9 means the indicator will alert wh...

Page 8: ...MODE enter counting mode and it displays n XXXXX Step 3 Press key to input the sample number and it displays n 00100 Step 4 Press MODE to confirm and upload the sample unit Step 5 Loading the weight s...

Page 9: ...ODE to confirm and enter into the next step Step 3 Display Resolution The screen displays XXXXX The value shown is the display resolution Display resolution division kg full capacity kg Example take 0...

Page 10: ...able 4 X 3 Indicator will send out 1 data after the weighing is stable Press to select the value and press MODE to confirm and finish to set the Serial protocol Step 6 Weight response speed Filtering...

Page 11: ...RROR SIGNAL There will shows error signal when the scale have some problem Error 1 means calibration isn t workable weight is too light or division is too high Error 2 means wrong zero check the load...

Reviews: