Sage 923lts User Manual Download Page 1

Sage 923LTS User’s Guide

Automated Local Loop Test

System

Sage Instruments

240 Airport Blvd.

Freedom, CA 95019

Tel. (831) 761-1000, Fax (831) 761-2452

www.sageinst.com

Part Number 9100-0923-01 v2.0

Advanced Test Equipment Rentals

www.atecorp.com  800-404-ATEC (2832)

®

Established 1981

Summary of Contents for 923lts

Page 1: ...ocal Loop Test System Sage Instruments 240 Airport Blvd Freedom CA 95019 Tel 831 761 1000 Fax 831 761 2452 www sageinst com Part Number 9100 0923 01 v2 0 Advanced Test Equipment Rentals www atecorp co...

Page 2: ...ocopied stored in a retrieval system distributed translated or transmitted in any form or by any means electronic mechanical recording or otherwise without the express written permission of Sage Instr...

Page 3: ...y 10 Special Screen Display Characters 10 Arrows 10 Blinking Cursor 10 Bracketed Selection 10 Using the Keypad 11 F1 and F2 11 Light Bulb Key 11 t u Backspace Delete and Forward 11 pq Up Down and Esca...

Page 4: ...uites 22 Automatic Test Updates 22 Manually Updating Tests 22 Using the 923LTS as a Telephone 22 Section 4 Technical Support 25 923LTS Technical Support 25 Internet Support 25 Frequently Asked Questio...

Page 5: ...confirm proper line installation and to monitor ongoing performance ALTS has two components 1 the Sage 923LTS hand held unit used in the field 2 a Sage 950RTS or 356Eplus responder located in the cen...

Page 6: ...to Noise Signal to Total Distortion Intermodulation Distortion Perceptual Speech Quality Measurement PSQM Other Tests The 923LTS may come equipped with other tests that do not require an ALTS responde...

Page 7: ...ontains instructions for the test sequence and each test s pass fail parameters The selected test runs automatically At the conclusion of the test the 923LTS displays an immediate overall pass fail no...

Page 8: ...8 9100 0923 01 V2 0 SAGE INSTRUMENTS...

Page 9: ...only the AC DC transformer supplied with the 923LTS to charge the batteries Other transformers may damage the 923LTS and void the warranty Initial Charge The 923LTS comes with rechargeable batteries i...

Page 10: ...ial Screen Display Characters Arrows An up or down arrow in the upper right corner of the display indicates that additional selections or data are available on other screens The number next to the arr...

Page 11: ...quence pq Up Down and Escape Use the p Up andq Down keys to scroll through lists that span several screens or to move to the previous or next screen You can use thep andq keys whenever an up or down a...

Page 12: ...ctions and can be used for both alphabetic and numeric data entry Use the keypad for making menu selections entering telephone numbers entering other information e g your identification number names f...

Page 13: ...e some of the basic operating features of the 923LT including Interfaces line types and ports Environment light and sound levels Phone lists responder phone numbers Interfaces Use the Interfaces scree...

Page 14: ...der number even if no dial tone is detected The maximum wait allowed is 99 seconds the default is 40 seconds No Answer Time Out While the 923LTS dials the responder to begin a test the display readsCa...

Page 15: ...number to change or to display a new empty screen 2 Press F1 to place the cursor on the number field 3 Use the keypad to enter digits 4 Use the u key to enter a one second pause in the dialing sequenc...

Page 16: ...sponder all available templates are automatically downloaded To download loop test templates from a responder to a new 923LTS follow these steps 1 From the Main Menu select Test to display the Test Se...

Page 17: ...e charged responder phone numbers are stored or available for entry test suites are available on the Sage 950RTS responder if required a service order number or trouble ticket number is available for...

Page 18: ...es a call from the responder back to the 923LTS If this screen is left blank then the call from the responder will not be performed 6 Press NEXT to display the technician identification screen 7 Enter...

Page 19: ...onder i e the responder that calls either the 100 test line or the 102 test line 7 Press NEXT to start the test 8 When testing is complete the 923LTS immediately displays complete detailed results NOT...

Page 20: ...splays the date and time in 24 hour notation of test calls 2 Use the keypad to select the desired call Use the p and q keys to scroll to more screens Up to ten calls can be stored on the 923LTS 3 Use...

Page 21: ...o the 923LTS including ring voltage ring frequency NOTE The call back option is applicable only if the test included a call initiated by the 950RTS responder in the Central Office to the 923LTS in the...

Page 22: ...o the Update Tests option 3 Use the q or p keys to scroll to the desired responder NOTE To download test suites from a new responder that does not appear on the display enter the phone number for the...

Page 23: ...or press NEXT to dial the displayed number 4 Dial access numbers and area codes just as you would from a regular telephone on the line 5 To hang up press TALK The 923LTS reverts back to a automated lo...

Page 24: ...24 9100 0923 01 V2 0 SAGE INSTRUMENTS...

Page 25: ...e link to access electronic forms to submit your request Frequently Asked Questions Why does the display light go out The display light automatically turns itself off to conserve power Press the backl...

Page 26: ...g both the CLR and the ON OFF keys until two lines of bars appear in the display 4 When the bars appear release the ON OFF key Continue to hold down the CLR key until the display goes blank 5 When the...

Page 27: ...rface Loop 600 ohm Serial Port 9600 bps 8 Data Bits No Parity Environment Display Contrast Dark Max Wait Dial Tone 40 seconds No Answer Timeout 40 seconds Backlight Timeout 15 seconds Audible Alert en...

Page 28: ...Sage Instruments is not liable for consequential damages or damages to any party other than the original buyer Sage Instruments specifically disclaims any implied warranty of merchantability or fitnes...