40
Model name
Classification according
to the degree of protection
against ingress of water (IEC 60529)
TRAUS FUS10
IPX8
6. Foot controller
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Electromagnetic Emissions
Electromagnetic environment - guide
Emission test
Compliance
The TRAUS SSG10 is intended for use in the electromagnetic environment specified below. The customer or the user of
the TRAUS SSG10 should assure that is used in such an environment.
RF emission – CISPR 11
EN 55011
Group 1
Class A
Class A
Compliance
Group 1 TRAUS SSG10 uses RF energy for internal operation. Therefore, its radiofrequency
emissions are very low and are not likely to cause any interference in nearby equipment.
TRAUS SSG10 is suitable to use in all establishments, including domestic
establishments and those directly connected to the public low voltage power
supply network that supplies buildings used for domestic purposes.
RF emission – CISPR 11
EN 55011
Harmonic emission –
IEC 61000-3-2
EN 61000-3-2
Voltage fluctuations
Flicker emission –
IEC 61000-3-3
EN 61000-3-3
In IEC/EN 60601-1-2. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
This equipment generate, use and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
- Consult the manufacturer or filed service technician for help.
Electromagnetic Immunity
Compliance level
Electromagnetic environment – guidance
Immunity test
IEC 60601 test level
The TRAUS SSG10 is intended for use in the electromagnetic environment specified below. The customer or the user
of the TRAUS SSG10 should assure that is used in such an environment.
Electrostatic discharge (ESD)
IEC 61000-4-2
EN 61000-4-2
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output lines
±2kV for power supply lines
±1kV for input/output lines
±1kV differential mode ±2kV
common mode
±1kV differential mode
±2kV common mode
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Electrical fast transients
IEC 61000-4-4
EN 61000-4-4
Surges
IEC 61000-4-5
EN 61000-4-5
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
EN 61000-4-8
NOTE:
UT is the a.c. mains voltage prior to application of the test level
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Voltage dips,
short interruptions and
voltage variations
IEC 61000-4-11
EN 61000-4-11
<5% UT (>95% dip in UT) for
0.5 cycle
40% UT (60% dip in UT) for
6 cycles
70% UT (30% dip in UT) for
30 cycles
<5% UT (>95% dip in UT)
for 5s
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 6 cycles
70% UT (30% dip in UT)
for 30 cycles
<5% UT (>95% dip in UT)
for 5s)
Mains power quality should be that of a
typical commercial or hospital environment.
If the user requires continued operation
during power mains interruptions, it is
recommended that the TRAUS SSG10 got
power from an uninterruptible power
supply or battery.
Summary of Contents for TRAUS SSG10
Page 1: ...www saeshin com USER MANUAL SSG10...
Page 2: ...SSG10...
Page 6: ...AC 100 120V 220 240V 3...
Page 8: ...LCD EMC EMC Electromagnetic Compatibility ISO 1797 1 1 2 3 4 1 2 3 4 5...
Page 14: ...SETUP SETUP SETUP 1 2 11 LCD...
Page 20: ...1 1 1 1 2 2 1 1 2 3 1 1 0 4 3 2 1 4 2 3 5 1 4 17...
Page 22: ...1 2 3 O X 19 4 A B C B 1 D C 1...
Page 26: ...23 DRILL SAW SHAVER 8 9 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3...
Page 36: ...33 M1 M2 M1 M2 1 10 20 3 10 3 3 70 30 2 1...
Page 42: ...IEC 60529 TRAUS FUS10 IPX8 6 39 1 1 2 SETUP SETUP SETUP EU CE BF 1639 NON STERILE...
Page 46: ...43 MEMO...
Page 47: ...44 MEMO...