34
Error code
Cause & Remedy
Cause: The foot controller connection is not proper or the foot controller is defective.
Remedy: Make sure that the foot controller is properly connected.
Contact the manufacturer if the remedies do not activated
Cause: The motor connection is not
proper or the motor is defective.
Remedy:
M
ake sure that motor properly
connected. Contact the
manufacturer if the remedies
do not activated.
Cause: Torque is overloaded than the setting values during the surgery.
Remedy: Press the speed control pedal to release the error.
Contact the manufacturer if the remedies do not activated.
Cause: Select motor without motor connector
(Motor 1 / Motor 2)
Remed
y: Press the motor button to be used (Motor 1 /
Motor 2). Contact the manufacturer if the
remedies do not activated.
■ Error & Remedy
: In case of stopping handpiece by not working & using error, each error code may be shown below.
1) Angle Handpeice
① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains
(blood, physiologic saline solution, etc.) and rinse.
※ Do not put tap water in the motor.
② Disconnect the angle handpiece to the motor.
③ Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes in Enzymatic Detergent.
④ Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
⑥ Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly subjected 3 times.
(Change the purified water each time.)
⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.
2) Motor
Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.
※ Pay attention to not enter the water into the motor.
■ Fuse replacement
1. Cleaning
Check the fuse in case the control box is malfunctioning.
Contact the manufacturer for replacement.
Fuse box
MAINTENANCE OF THE PRODUCT
Summary of Contents for TRAUS SSG10
Page 1: ...www saeshin com USER MANUAL SSG10...
Page 2: ...SSG10...
Page 6: ...AC 100 120V 220 240V 3...
Page 8: ...LCD EMC EMC Electromagnetic Compatibility ISO 1797 1 1 2 3 4 1 2 3 4 5...
Page 14: ...SETUP SETUP SETUP 1 2 11 LCD...
Page 20: ...1 1 1 1 2 2 1 1 2 3 1 1 0 4 3 2 1 4 2 3 5 1 4 17...
Page 22: ...1 2 3 O X 19 4 A B C B 1 D C 1...
Page 26: ...23 DRILL SAW SHAVER 8 9 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3...
Page 36: ...33 M1 M2 M1 M2 1 10 20 3 10 3 3 70 30 2 1...
Page 42: ...IEC 60529 TRAUS FUS10 IPX8 6 39 1 1 2 SETUP SETUP SETUP EU CE BF 1639 NON STERILE...
Page 46: ...43 MEMO...
Page 47: ...44 MEMO...