SABLE SLOW SPEED ATTACHMENTS
MAINTENANCE AND CLEANING MANUAL
Please retain this manual as your proof of purchase.
MODEL No. DESCRIPTION
Head Maximum RPM (5K or 20K)
1600201
1:1 NSK Type Contra -Angle
1600401 Snap on Prophy Head (5K)
1600202
4:1 NSK Type Contra -Angle
1600402 Screw-in Prophy Head (5K)
1600203
16:1 NSK Type Contra -Angle
1600403 Standard latch Head (5K)
1600204
1:1 MW Type Contra Angle
1600407 BB swing latch Head (20K)
1600210
1:1 NSK Type Contra -Angle
1600405 Miniature latch Head (5K)
1600301
1:1 Straight Nosecone
1600404 BB push Button Head(20K)
1600302
1:1 MW Shorty &
1600411 BB BP Latch Head (20K)
Rhino Type Nosecone
1600413 Sable friction grip Head (5K)
1600303
4:1 Straight Nosecone
1600412 StarType BB PB latch head (20K)
1600304
1:1 Lab Nosecone
1600406 StarType 12 tooth Standard latch
head(5K)
1600310
1:1 Straight Nosecone
1600408 StarType snap-on head (5K)
1600501
Midwest type snapon head (5K)
1600409 Star type screw-in head(5K)
1600502
Midwest type screwin head (5K
1600410 Star type BB SL head (20K)
1600503
Midwest Standard latch head (5K) 1600504 Midwest BB PL head (5K)
SAFETY PRECAUTIONS:
1. Always exercise extreme caution when operating your Slowspeed Attachment
2. Always inspect your Slowspeed Attachment thoroughly before each use.
3. Ensure proper burs are used for specific product. Use only burs that are DIN EN ISO 1797-1
norm and are made from hardened steel.
4. Do not use burs that are bent, worn or show any signs of damaged.
5. Ensure that the proper speed(rpm) attachment is used when selecting attachment and motor for Heads
6. Always use PPE (Proper Protective Equipment) when operating and cleaning your Slowspeed
Attachment See your Occupational Health and Safety Act or other safety ordinances for specific
instructions.
7. If product is dropped your dealer should be contacted, it is recommended that you do not use.
8. Contact your dealer right away if any problem should occur with any of your Sable products.
9. Sterilize before first use and before each subsequent use there after
10. This product is for use by a dentist or trained professional in the use of this product.
FEATURES
1. Attachments conform to International Standard Organization (ISO) specifications.
2. Will attach to all E type (ISO) Air Motors.
3. Autoclavable.
SPECIFICATIONS
1. Connection with Air Motor:
ISO Standard
2. Method of Sterilization:
Autoclave
HOW TO USE
1. Slide Attachment onto shaft of E Type or MW type Air Motor until locked in place.
2. To remove Attachment: pull Attachment, pushing release button located on the exterior of
Air Motor.
3. Direction of operation is determined by Air Motor.
LUBRICATION
USE ONLY AUTOCLAVABLE HANDPIECE LUBRICANT.
1. Remove Attachment from Air Motor.
2. Remove Head from Attachment.
3. Apply lubricant into both ends of Attachment and reapply to Air Motor.
4. Run Air Motor and Attachment at normal speed for 10 seconds.
5. Wipe off excess oil from Attachment surface.
AUTOCLAVING PROCEDURE
1. Wipe off debris on the Attachment surface with alcohol-soaked cloth.
2. Lubricate the Attachment. (Refer to Lubrication Instructions).
3. Autoclave for 20 minutes at 135 C.
ATTENTION
This is a precision instrument. Do not drop or apply excessive force when handling.
WARRANTY INFORMATION
The Sable slow speed Attachment is warranted to be free of manufacturing defects for a
period of 3 months* from date of purchase.
*Warranty is conditional on proper use and maintenance of Air Motor. Warranty is limited
to repair or replacement at manufacturer’s discretion. Under no circumstances will the
manufacturer be held responsible for incremental or consequential damages.
Please contact point of purchase authorized distributor for warranty repair.