Setting up
Mounting, fastening
With the table mounting delivered, the mikron 81 detector can be operated in a laboratory,
but can also be inserted into a device insofar as the ambient conditions (temperature,
humidity, composition of the atmosphere) are maintained.
When inserting the apparatus into a device, fastening it at the front of the detector block is
recommended (see Fig. 4), as this would not constrict the cross-section and otherwise the
housing would be deformed. A suitable means of fastening is, for example, pipe clips
conforming to DIN 3015.
As liquid only flows through the measurement cell, a leak can occur in the cell. To avoid
damage to electronic parts of the detector through the presence of liquid, it is recom-
mended when inserting the measurement cell vertically or at an angle to ensure that its end
cap is located at the deepest point. In the case of horizontal insertion the longitudinal
alignment groove should be situated at the bottom along with the overflow outlet of the