1. Production Introduction

1.1 Production Synopsis

DPS3012/DPS5015/DPS5020 combines analog adjustment and digital control, at the

same time it is compact and powerful, and becomes the first choice for many enthusiasts and
laboratory workers.

The unit offers non-volatile power off parameter storage and 10

programmable preset memory Data Groups with a shortcut recall function for groups 1 and 2.
All these productions are designed with LCD display, so we can check set Voltage, set Current,
output Voltage, output current, output power, input Voltage and so on. On the right side of
the main interface, we can see output enabled or disabled, Constant Current or Constant
Voltage mode of operation, abnormal condition warning, keypad locked or unlocked state
and the Data Group in use. The data setting interface allows adjustment of Over-Current,
Over-Voltage and Over-Power protection values, Preset value, LCD brightness and whether
the power is turned on or the like.

DPH5005 has 3 versions, Non-communication version, USB communication version, USB

+ Bluetooth communication version. Communication version can be connected to PC,
Bluetooth version can be connected to Android mobile phone.

(Note: DPH3205 only has no

communication)

This module has many advantages, small size, advanced function, good visual effect,

high operability, high-precision, it can be used independently or be inset into the device and
been widely applied.

1.2 Series Product Parameters

产品型号

DPH5005

DPH3205

Input Voltage range

DC6-50V

DC6-40V

Output Voltage range

0-50.00V

0-32.00V

Output power range

0-250W

0-160W

Output Current range

0-5.000A

0-5.000A

Summary of Contents for DPH Series

Page 1: ...so we can check set Voltage set Current output Voltage output current output power input Voltage and so on On the right side of the main interface we can see output enabled or disabled Constant Curre...

Page 2: ...0mV VPP connection cables length 10 45 Note the following description uses the DPH5005 as an example Other models have same functions and operation except for parameters 1 3 Connection Description IN...

Page 3: ...rse will cause damage input over current protection cannot be triggered frequently because that will cause damage to the model so be careful when using the module 1 4 Panel Description A Voltage setti...

Page 4: ...1 6 2 Advanced Parameter Setting On the main interface press the key to enter the Data Setting Interface In the Data Setting Interface press key or key to step up or down through the menu options to s...

Page 5: ...It s up to you to set the brightness you like 1 6 2 3 Data Settings and Saving in Specified Data Group Switch to press the rotary control to select Data Group Number press key and then press key or k...

Page 6: ...M0 M9 The M0 Data Group is the default Data Group after power on It should be noted that any time an alternative Data Group is selected or any changes made to the existing settings then M0 will be im...

Page 7: ...ter through the Micro USB cable Right click MY COMPUTER then click Manage Device Manager and click the drop down arrow at Port to see the module s COM port number as shown below Note If the USB commun...

Page 8: ...nitialize the communication parameters of the Bluetooth module if it is the first time to use Bluetooth communication function The red LED of Bluetooth communication module starts to flash after the d...

Page 9: ...l port of the Bluetooth connection Note Win 7 system select the Bluetooth device right click to view the properties display the following window view the COM port of the Bluetooth settings select the...

Page 10: ...ommunication data of digital power supply you must disconnect Bluetooth with PC the LED will be flashing Otherwise you can change the data of Bluetooth board For computer operating system please insta...

Page 11: ...27 32 Double click the Arial Unicode MS Font Pack and click Install to install the Font Pack 2 4 2 Double click setup exe to Set Up Computer Software...

Page 12: ...tion 2 3 for Bluetooth connection section 2 1 for USB connection baud rate slave address default 9600 001 click On Line to start communication and automatically lock the digital power button after suc...

Page 13: ...ing shortcut recall brightness adjustment voltage and current curve derivation You can set Voltage Current by turning the knob or enter a value The graph on the right shows the real time voltage and c...

Page 14: ...oogle Store can also download the file with the extension APK and install it on the phone You can get download address in section 2 4 3 1 2 Application Setup Android APP can only work in Android 5 0 o...

Page 15: ...e Bluetooth on the phone first then turn on the app Bluetooth connection to connect to the Bluetooth directly AD program information Version number description assistance developer information AE Volt...

Page 16: ...32 32 4 Others APP for IOS has not been finished You can send E Mail to 1749808860 qq com if you have any question...

Reviews: