Documenting Multifunction Calibrator / EN
06/04/2018 Version No. 1
35
2. Current Scanning for SOURCE
2a. How to setup:
(1) Press
button to enter SETUP function.
(2) mA 0%: set up the
“starting” current for scanning.
(3) mA 100%: set up the
“ending” current for scanning.
Remark:
1.
Users are allowed to type in max. 5 digits.
2.
Type in a current value (including decimal point), press ENTER, then
the calibrator will output this current value.
3.
When the output value is <0, please type in a negative sign first.
4.
When the output value is <1 and >0,
please type in “0.” first.