Documenting Multifunction Calibrator / EN
06/04/2018 Version No. 1
23
Remark:
1.
Users are allowed to type in max. 5 digits.
2.
Type in each parameter (including decimal point), press ENTER, then the
calibrator will output the parameter values.
3.
When the output value is <0, please type in a negative sign first.
4.
When the output value is <1 and >0,
please type in “0.” first.
Warning:
1.
Do not make an input or connect any charged installment to SOURCE
(terminals) to prevent from damaging the calibrator.
2.
When there is an output short circuit or overload, the calibrator can not
output the correct frequency/waveform.
3.
Perform one function at a time and make connection to the specific
terminals only. Remove all other connections to the unused terminals.
Always connect to only one of SOURCE , MEASURE , or TC alone.
6b. How to set up
(1) Hz: set up the output frequency of F1 and F2.
(2) %: set up the output % of F1 and F2.
(3) Phase: set up the outset phase angle of F1 and F2.
(4) Vpp: set up the output Peak to Peak voltage.
(5) Offset: set up the output DC bias accurate position.