Documenting Multifunction Calibrator / EN
20
06/04/2018 Version No. 1
5. Pulse Function
5a. 3.0μS~999.99mS , 0.5Vpp~10Vpp , offset:-5V ~ +5V
(1) Turn off the power. Sliding switch turned to Hz.
(2) Press
button and
button at the same time; the display
will show a reverse
“PULSE” mark which means it already enters into
the PULSE function.
(3) Press
button to switch among
“Voltage Peak to Peak (Vpp)”,
“OFFSET” and “PULSE”.
(4) Then type in a value (including decimal point) and press ENTER.
(5) Connect Test leads or Alligator clips to SOURCE terminals (red to +, and
black to -).
(6) Connect Test leads or Alligator clips to the object for calibration.