ABX open
•
ABX closed
•
Roche (can be used with software version 5.02 and up)
See for rack example information
Make your selection, press
Save settings
.
6.7.1.3.4 - Test barcode reader
The following steps are needed for checking the barcode.
Take a sample tube with barcode and a rack.
1. Go to the
IN position
.
2. Go to
Mix position
.
3. Go to
Tube position number
1.
After pressing Read barcode
the barcode is shown above the button. If barcode label is not read correctly
see section Troubleshooting.
6.7.1.3.5 - Test tube sensor
Switch
Check tube detection sensor
:
Use this switch to check the correct adjustment of the optical sensor that checks for the presence of sample
tubes in the rack. See “
Troubleshooting/ Tube sensor does not detect tubes
” for further instructions.
Switch
Always read barcode
:
Use this switch to read the barcode at every position in the rack. This tube is correctly logged in the rack,
even when the tube sensor is not detecting a tube. If the tube sensor is not detecting a tube and a barcode
could also not be read, the tube position is logged as empty.
See Service manual for further instructions
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
99
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...