▪ Select Set rinse position.
▪ Change Rinse position.
▪ Enter the correct value of the pipette number at the rinse station.
▪ Close pop-up menu.
2. Select Set rinse position
▪ Change Rinse position
▪ Press Learn carousel position
▪ When all positions are learned, close pop-up menu.
Check if the position error is solved: Select
Go to rinse position
, enter a higher value than the actual rinse
position and the carousel should move without position error.
If position error is solved:
1. Leave service mode.
2. If samples or liquid is present in any pipette, go to tab
Maintenance - Prime/clean - Wash all pipettes
which
removes the liquid from the pipettes.
All sample information will be lost.
If there is still a position error:
If the position is still not correct or if position errors occur frequently and the solutions above do not solve
the problem, the following procedure must be carried out:
1. Leave Service mode
2. a) If liquid or samples are present in isolated pipettes, use a combination of
Go to rinse position
at the
service tab and
Maintenance -Prime rinse solution
to remove the blood samples. All sample information will
be lost.
or
b) Go to tab
Settings
and select
Delete pipette data
. The carousel stops moving. Use the sequence
mentioned above to remove the blood samples.
3. Check mechanical connection potentiometers.
If the position error is not solved after all the mentioned actions, the position device needs replacement.
This has to be done by a trained technician, call distributor.
15.15 - Separator error
If it takes too long for the waste pump to empty the waste separator, the system generates a separator
error.
Separator error may be caused by:
Extensive foam build-up in the waste
separator.
Check the separator assembly and connections for possible
air leaks.
Waste-tube between waste separator and
waste pump is blocked.
Replace the tube.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
161
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...