15.10 - Rinse nozzle (wash station) alignment errors
If there is a mechanical obstruction, the rinse nozzle may not align correctly with the pipette. Check that the
two pieces of plastic tubing connected to the rinse nozzle have enough slack to allow for movement of the
nozzle.
Technical support is needed if the rinse nozzle does not line up with the pipette.
15.11 - Hazy reports
"Hazy" reports are usually caused by build-up of proteins on the inner wall
of the pipettes. Another cause is growth of micro organisms in the diluter
system. It is extremely important that the system is kept sterile.
Erroneous haziness looks more like haemolytic plasma than 'normal' hazy
plasma. If there is an abnormal number of 'hazy' aspects, check the
dispenser system carefully and clean it if there is any doubt.
Symptoms for a contaminated instrument:
1. If more than
3… 5
out of
20
measurements are reported
HAZY.
2. Poor meniscus may be reported.
3. Incorrect results.
In case of hazy reports:
1. First run an extra Fill and Clean sequence.
2. Check after a day's run if haziness is decreased. See
3. When there are still many reports, it is recommended to clean the
diluter system. See
WI-240 Cleaning the diluent system
Example of haziness
Troubleshooting
158
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...