7.3.1 - Starrsed Rinse solution QRR 010934
Pre flush reagent bottle with de-ionized water.
Fill rinse container with rinse solution from the 20-liter container.
7.3.2 - Starrsed Saline QRR 010933
Pre flush the saline bottle using saline from the saline container. Fill saline bottle from the 5-liter container.
7.3.3 - Starrsed Diluent
Starrsed Diluent is a dedicated IVD product, developed exclusively for the automatic dilution of blood
samples in Starrsed ESR analyzers.
•
Approximately 0.5 ml Diluent is used for each sample.
•
Approximately 2.5 ml is used for one Prime cycle.
The diluent container is a 5-liter container (QRR 010931).
7.3.4 - De-ionised water
Add one or two drops of Starrsed Saline to the deionized water to avoid
Reagent alarm
.
7.3.5 - Starrsed Disinfectant QRR 010947
Pre flush the disinfectant bottle using disinfectant from the disinfectant container. Fill disinfectant bottle from
the 5-litre container.
7.3.6 - Starrsed Cleaning Agent
Starrsed Cleaning Agent is used to remove protein deposition from the Westergren pipettes of Starrsed ESR
analyzers.
Starrsed Cleaning Agent needs to be prepared for a cleaning procedure which is used in level 4
maintenance.
Starrsed Cleaning agent is supplied in 100 ml bottles.
QRR010905: 1 bag of 2 bottles
Getting started
MRN-071-2-EN version 7.02 Starrsed RS
111
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...