1.
Valve outer needle:
Energizes the outer needle solenoid valve.
2.
Sample control solenoid
: Sample control solenoid fill sequence energized. The function of this solenoid is
to build up a vacuum in the Westergren pipette before the aspiration cycle starts.
3.
V-Valve wash section
: Pipette wash vacuum control valve, controls the main vacuum line between the
wash-station and separator.
4.
V-Valve fill section
: Vacuum control fill-nozzle / sample probe, controls the main vacuum line between the
fill nozzle cap and separator.
Actuators
1.
Rinse actuator
: Rinse solenoid active, rinse valve-block down.
2.
Fill actuator
: Fill solenoid active, fill valve-block down.
6.7.5 - Display error history (Service)
•
Load settings
is used for reloading the stored software settings.
•
Save settings as
is used for storing software settings to a file. There are no restrictions for the file
name.
•
Save settings
is used for storing software settings after settings are changed or altered.
A list of the error detection during operation.
This list contain error codes that can help troubleshooting the instrument.
The key
Clear
will delete all errors from the list (not advised option).
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
107
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...