background image

E-M-HP22-V2_12 

Rotronic AG 
Bassersdorf, Switzerland 

Document code 

Unit 

HygroPalm HP22-A hand-held indicator: 

User Guide 

Instruction Manual 

Document Type 

Page 

1

 of 16 

Document title 

 
 

© 2010-2012; Rotronic AG                                                                                                                             E-M-HP22-V2_12

 

 

HygroPalm HP22-A Hand-Held Indicator 

User Guide 

 

 

 

 

 

Summary of Contents for HygroPalm HP22-A

Page 1: ...rsdorf Switzerland Document code Unit HygroPalm HP22 A hand held indicator User Guide Instruction Manual Document Type Page 1 of 16 Document title 2010 2012 Rotronic AG E M HP22 V2_12 HygroPalm HP22 A...

Page 2: ...ction between the HP22 and probe functions 7 4 Operation 8 4 1 Display 8 4 2 Keypad 8 4 4 ON OFF key 9 4 5 Changing the parameters shown on the display 9 4 6 HOLD function 9 4 7 Changing the unit syst...

Page 3: ...0 to 100 RH and 100 to 200 C 148 to 392 F The temperature operating range of the HP22 electronics is limited to 10 60 C Based on the ROTRONIC AirChip 3000 digital technology the HP22 indicator and Hy...

Page 4: ...n the HP22 example HP23 hand held calibrator This allows showing the data measured by the HP22 on the other instrument display and also allows using the other instrument internal menu to do for exampl...

Page 5: ...ed data can be downloaded to a PC with the HW4 software time stamped and viewed MEASUREMENT LOOP VALIDATION Functions Description Simulator mode Used to make the HP22 generate fixed values for the hum...

Page 6: ...e adjustment K Device write protection Disabled HP22 and probe Menu access from keypad Enabled Limit humidity output to 100 RH Enabled probe Out of limit value digital display alarm Disabled HP22 and...

Page 7: ...ce name of the HC2 probe is not displayed by HW4 and is replaced with the HP22 Input Name Calculation X X Psychrometric calculation HP22 setting overrides HC2 probe setting Data refresh rate X When se...

Page 8: ...t point and the bottom line corresponds to temperature The display can be configured to show a trend indicator on each line increasing value decreasing value In the event of an alarm the display shows...

Page 9: ...alculated at the time the ON OFF key was pressed Likewise communication with the ROTRONIC HW4 software is frozen The ENTER key can still be used to change the parameters shown on the display but the v...

Page 10: ...Firmware version Serial Nbr Serial number Address RS 485 address Name Device name User defined SensorTest Humidity sensor status Off Good SQ Tuned Bad Record On Off Data recording by the probe max 200...

Page 11: ...ons condensation may occur on the sensors when the probe is colder than the environment As long as the humidity temperature limits of the humidity sensor are not exceeded condensation does not alter t...

Page 12: ...int adjustment of temperature The effect of a 1 point temperature adjustment is to add the same offset to all measured values When temperature is stable press the MENU key to show the internal menu on...

Page 13: ...points are automatically deleted from the probe memory after an adjustment o Instructions for using the ROTRONIC calibration devices and humidity standards are provided in document E M CalBasics o Bec...

Page 14: ...chnical data 7 1 Specifications General HP22 Device type Humidity temperature hand held indicator with one probe input HygroClip 2 Battery type 9 V alkaline Low battery indication Yes Humidity and tem...

Page 15: ...g Temperature limits at probe Depends on probe model Maximum humidity at probe 100 RH up to 80 C 176 F 75 RH at 100 C 212 F 45 RH at 125 C 260 F 15 RH at 150 C 302 F Maximum air velocity at probe 20 m...

Page 16: ...3 Probe Adjustment function AirChip 3000 devices E M HW4v3 DR 001 HW4 software version 3 Data Recording Function AirChip 3000 Devices E M AC3000 CP AirChip 3000 Communication Protocol E M CalBasics Te...

Reviews: