Manufacturer: Rossmax International Ltd.,

12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan

www.rossmax.com

 Specifications

 

 Cautions and maintenance

1. This device should be used with nebulizer only.

2. Place the device and nebulizer on a flat, horizontal surface while in use.

3. The device must be in a vertical position while in use.
4. Turn the regulator slowly to avoid possible damage.
5. Do not completely unscrew regulator stem. 
6. Do not attempt to disassemble the device.

7. Avoid excessive vibration, physical impact and dropping.
8. Do not store the device in direct sunlight, high temperature or humidity.

9. Keep the device away from static electricity.

10.  For the first use, open the rubber vent-plug on the back side of pressure gauge and 

then close it in order to balance the pressure after the transportation.

Note:  Do not open the rubber vent-plug when the device is in a vertical position to avoid 

leakages of oil material from the pressure gauge.

Power source

No need

Operating temperature

0 ~ 54ºC

Storage temperature

- 10 ~ 60ºC

Air flow

scale

1 ~ 10 L/min

accuracy

± 4% of full scale

Pressure

scale

0 ~ 60 PSI

accuracy

± 2% of full scale

Weight

1000 g

Dimension

128 (L)*120(W)*158 (H) mm