Rossmax AD761 Instruction Manual Download Page 1

1. Connect the AC adapter with the AC adapter jack in the back 

of the unit.

2. Plug the AC adapter into the socket. (AC adapters with re-

quired voltage and current indicated near the AC adapter 
jack.)

Caution:

 1.  Please unload the batteries when operating with 

the AC mode for a longer period of time . Leav-
ing the batteries in the compartment for 
a long time may cause leakage, which 
may lead to damage of the unit.

                2.  No batteries are needed when 
                   operating with the AC mode.
                3.  AC adapters are optional. Please contact 

the distributor for the compatible AC adapters.

                4.  Use  only  the  authorized  AC  Adaptor  with  this  blood  pressure 

monitor. 

12.  Using the AC Adapter (Optional)

AC Adapter

(Ø3.8/Ø1.3)

13.  Installing Batteries

1.  Press down and lift the battery cover in the direction of the arrow to open 

the battery compartment.

2.  Install or replace 4 “AA” sized batteries in the battery compartment accord-

ing to the indications inside the compartment.

3.  Replace the battery cover by clicking in the bottom hooks first, then push in 

the top end of the battery cover.

4.  Replace the batteries in pairs. Remove batteries when unit is not in use for 

extended periods of time.

You need to replace the batteries when

              1.  low battery icon appears on display.
              2.  the START/STOP key is pressed and nothing appears on display.

Caution

1.  Batteries are hazardous waste. Do not dispose them together with the 

household garbage.

2.  There are no user serviceable parts inside. Batteries or damage from old 

batteries are not covered by warranty.

3.  Use exclusively brand batteries. Always replace with new batteries together. 

Use batteries of the same brand and same type. 

1.  Unwrap the arm cuff, leaving the end of the cuff through the D-ring of 

the cuff.

2.  Put your left arm through the cuff loop. The color strip indication should be 

positioned closer to you with the tube pointing in the direction of your arm 
(Fig. 1). Turn your left palm upward and place the edge of the arm cuff at 
approximately 1.5 to 2.5 cm above the inner side of the elbow joint (Fig. 2). 
Tighten the cuff by pulling the end of the cuff.

14.  Applying the Cuff

The 

National High Blood Pressure Education Program Coordinating 

Committee

 has developed a blood pressure standard, classifying blood 

pressure ranges into 4 stages. 

(Ref. The Seventh Report of the Joint Na-

tional Committee on Prevention, Detection, 
Evaluation, and Treatment of High Blood 
Pressure-Complete Report JNC-7, 2004).

This blood pressure classification are based 
on historical data, and may not be directly 
applicable to any particular patient. It is 
important that you consult with your physi-
cian regularly.  Your physician will tell you 
your normal blood pressure range as well 
as the point at which you will be considered 
at risk. For reliable monitoring and refer-
ence of blood pressure, keeping long- term 
records is recommended. Please download 
the blood pressure log at our website www.
rossmaxhealth.com.

9.  Display Explanations

8. 

 Important Notes: Regarding Your Blood Pressure Measurement

Blood pressure fluctu

-

ates all the time!

You should not be overly 
worried if you encoun-
tered two or three meas-
urements at high levels. 
Blood pressure changes 
over the month and even 
throughout the day. It is 

also influenced by season 

and temperature.

Rossmax HSD (Hemodynamic Stability) technology communicates to a user 
if a measurement is taken in the appropriate state of rest ensuring more 
correct blood pressure readings. The most common error in blood pressure 
measurement is taking the measurement when the users not being at rest 
physically or mentally or hemodynamically unstable thus measurements 
can be affected and incorrect.  

Physical and mental factors can have a significant impact on actual blood 

pressure.  Simply walking up stairs has a measurable affect on blood pres-
sure.  By monitoring changes of the heart rate frequency throughout a 
blood pressure measurement, Rossmax HSD can determine if a diagnosis is 
taken in the appropriate physiological state of rest.  Rossmax HSD prompts 
the user to disregard results taken when not at rest and retake another 

5.  Blood Pressure Standard

6.  Blood Pressure Fluctuation

7.  Resting Detector Using HSD Technology

The National High Blood Pressure Educa-
tion Program Coordinating Committee 
has developed a blood pressure standard, 
classifying blood pressure ranges into 4 
stages. This unit is equipped with innova-
tive blood pressure risk indicator, which 
visually indicates the assumed risk level 
(prehypertension / stage 1 hypertension / 
stage 2 hypertension) of the result after 
each measurement. 

This unit is equipped with an Irregular Heartbeat 
Detector (IHB) which allows those who have an ir-
regular heartbeat to obtain accurate measurements 
alerting the user of the presence of an irregular 
heart beat during the measurement.

Note:

  It is strongly recommended that you consult 

your physician if the IHB icon ( 

 ) appears 

often.

10.  Hypertension Risk Indicator

11.  Irregular Heartbeat Detector (IHB)

1.

 Date/Time Indicator

2.

 Hypertension Risk Indicator

3.

 Memory Zones

4.

 Resting Detector

5.

 Memory Average

6.

 Memory Sequence Number

7.

 Weak Battery Mark

8.

 Irregular Heartbeat Detector (IHB)

9.

 Systolic Pressure

10. 

Diastolic Pressure

11. 

Pulse Rate

12. 

Pulse Mark

1.  Take your reading in a comfortable environment as measurements can 

be affected by hot or cold temperatures. Take your blood pressure at 

normal body temperature.

Attention:

  Consult the accompanying documents. Please read this manual 

carefully before use. For specific information on your own blood 

pressure, contact your physician.  Please be sure to keep this 
manual.

Blood pressure measurements determined with AD761f are equivalent to 
those obtained by a trained observer using cuff/stethoscope auscultation 
method, within the limits prescribed by the American National Standard, Elec-
tronic or Automated Sphygmomanometers. This unit is to be used by adult 
consumers in a home environment. Do not use this device on infants or neo-
nates. AD761f is protected against manufacturing defects by an established 
International Warranty Program. For warranty information, you can contact 
the manufacturer, Rossmax International Ltd.

1.  Introduction

3.   Real Fuzzy Measuring Technology

4.  Preliminary Remarks

This unit uses the oscillometric method to detect your blood pressure. Be-

fore the cuff starts inflating, the device will establish a baseline cuff pres

-

sure equivalent to the air pressure. This unit will determine the appropriate 

inflation level based on pressure oscillations, followed by cuff deflation.
During  the  deflation,  the  device  will  detect  the  amplitude  and  slope  of 

the pressure oscillations and thereby determine for you the systolic blood 
pressure, diastolic blood pressure, and pulse.

This Blood Pressure Monitor complies with the European regulations and 
bears the CE mark “CE 0366”. The quality of the device has been veri-

fied and conforms to the provisions of the EC council directive 93/42/EEC 
dated 14 June 1993 concerning medical devices:
EN 1060-1: 1995/A1: 2002 Non-invasive sphygmomanometers 

- Part 1 - General requirements 

EN 1060-3: 1997/A2: 2009 Non-invasive sphygmomanometers 

- Part 3 -  Supplementary requirements for electro-mechanical blood pres-

sure measuring systems

EN 1060-4: 2004 Non-invasive sphygmomanometers 
- Part 4:  Test Procedures to determine the overall system accuracy of auto

-

mated non-invasive sphygmomanometers.

This blood pressure monitor was designed for long service time. To ensure 
continued accuracy, it’s recommended that all digital blood pressure mon-
itors require re-calibration. This monitor (under normal usage with approx. 
3 measurements a day) does not require re-calibration for 2 years. Once 
the unit should be re-calibrated the device will display 

 . 

The unit should also be re-calibrated if the monitor sustains 
damage due to blunt force (such as dropping) or exposure 

to fluids and / or extreme hot or cold temperature / humid

-

ity changes. When 

 appears, simply return to your nearest 

dealer for re-calibration service.

 www.rossmaxhealth.com

2.  Name/Function of Each Part

Date/Time 

Set Key

Arm Cuff

LCD Display

User-Switching Key

START/STOP Key

Memory Key

Air Tube and 

Connector

Battery Cover 

(Located on back of unit)

4“AA”(R06) size,

1.5V batteries.

Data Link 

Socket

AC Adapter Jack

Resting 

Detector 

LED

measurement.
Blood pressure measurement can be affected not only by one’s physiologic 
conditions but also physical load, mental stress and certain disorders of 

the body:

Physical Causes

Actual Exercise
Post Exercise

Mental Causes

Excitement
Anxiety
White Coat Effect
Distraction

Talking during measurement

Medical Causes

Cardiac Arrhythmia
Impaired Vasal Control
Medication
Hyperventilation

Pulse Period & Pulse Pattern Changes during measurement 

Blood Pressure Unstable during measurement

HSD Instability Indication

 Hemodynamic Stability

indicates that the circulatory system is sufficiently at rest when taking the 

measurement thus measurement results are reliable and good for refer-
ence.

 Lack of Hemodynamic Stability

indicates  that  the  circulatory  system  is  not  sufficiently  at  rest  (hemody

-

namic instability) when taking the measurement thus the measurement 
results should be ignored. Wait a minimum of 5 minutes before re-taking 
measurements. More time may be needed depending on one’s physilogic 
conditions. 
If 

 icon repeatedly appears, see page 11 for important notes regarding 

your blood pressure measurements.  If 

 continues to appear even after 

long period of relaxation, this indicates the circulatory system is still not 

sufficiently  at  rest  and  cannot  be  relieved  within  a  short  period  of  rest. 

Possible causes are, but not limited to, inner unrest of the nervous system, 
chronic mental conditions or the developing cardiac arrhythmia that can 
cause hemodynamic instability for a 
long time and may persist continu-
ously even after long period of rest.
Rossmax HSD is simple to use and 
requires no additional user skill 
and device adjustments. Rossmax 
HSD uses only one measurement to 
determine if the users were at rest 
(HSD NEGATIVE) or not at rest (HSD 
POSITIVE) while taking the measure-
ment.

HSD Negative

(at resting state)

HSD Positive

(not at resting 

state)

2
3

7

1

9

10

11
12

8

5
6

4

2.  Do not move or talk during measurement as this can elevate readings.

3.  Do not move or cross legs during measurements. Keep feel flat on floor

4.  It’s suggested that you take your measurements at the same time each 

day and use the same arm for consistency.

5.  Users should wait a minimum of 5 minutes before taking additional 

measurements. More time may be needed depending on your physiol-

ogy.

6.  This product is not suitable for people with arryhthmias

7.  This device may have difficulty determining the proper blood pressure 

for users with diabetes, poor circulation of blood, kidney problems, or 

for users who have suffered a stroke.

Instruction 

Manual

Model: 

AD761  f

EN