IB-106-340
2-12
Do not use 100% nitrogen as a low gas
(zero gas). It is suggested that gas for the
low (zero) be between 0.4% and 2.0% O
2
.
Do not use gases with hydrocarbon con-
centrations of more than 40 parts per
million. Failure to use proper gases will
result in erroneous readings.
TO ELECTRONICS
REF AIR SET
263C152G01
1
FLOWMETER
0.2-2.0 SCFH 771B635H02
2
2" PRESSURE GAGE
0-15 PSIG
275431-006
3
COMBINATION FILTER-REG.
0-30 PSIG
4505C21G01
NOTE: DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
1
2
3
4.81 (122.17)
FLOW SET
POINT KNOB
0.125-27 NPT FEMALE
OUTLET CONNECTION
1.19
(30.22)
10.0
(254)
REF
DRAIN VALVE
3.12 (79.25) MAX
8.50
(215.90)
MAX
2.0
(50.80)
2 MOUNTING HOLES
3.19 (81.03) LG
THROUGH BODY FOR
0.312 (7.92) DIA BOLTS
1.50
(38.10)
2.250 (57.15)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER 4000 PROBE HEAD.
OUTLET
0.25-18 NPT FEMALE
INLET CONNECTION
INSTRUMENT AIR SUPPLY
10-225 PSIG MAX PRESSURE
26170035
0.250 OR 6 MM O.D. TUBING
(SUPPLIED BY CUSTOMER)
FITTING (SUPPLIED BY WECO)
TUBE COMPRESSION
0.250 OR 6 MM O.D.
500 VA
SERIAL NO.
TAG NO.
OXYMITTER 4000
WATTS:
VOLTS:
FUSE:
LINE
OUTPUT:
Rosemount Analytical Inc.
Orrville, OH 44667-0901
85-264 VAC 48-62 Hz
TM
800-433-6076
4-20 mA
R
5 Amps
TM
HART
SMART FAMILY
Figure 2-11. Air Set, Plant Air Connection
2-4.
PNEUMATIC INSTALLATION (FOR OXY-
MITTER 4000 WITHOUT SPS 4000).
a.
Reference Air Package. After the Oxymitter
4000 is installed, connect the reference air set to
the Oxymitter 4000. The reference air set should
be installed in accordance with Figure 2-11.
Instrument Air (Reference Air): 10 psig
(68.95 kPag) minimum, 225 psig (1551.38 kPag)
maximum at 2 scfh (56.6 L/hr) maximum; less
than 40 parts-per-million total hydrocarbons.
Regulator outlet pressure should be set at 5 psi
(35 kPa). Reference air can be supplied by the
reference air set of the IMPS 4000.
If using an IMPS 4000, refer to the IMPS 4000
Intelligent Multiprobe Test Gas Sequencer
Instruction Bulletin for the proper reference air
connections.
b.
Calibration Gas. Two calibration gas concen-
trations are used with the Oxymitter 4000, Low
Gas - 0.4% O
2
and High Gas - 8% O
2
. See Figure
2-12 for the Oxymitter 4000 connections.
500 VA
R
SERIAL NO.
TAG NO.
OXYMITTER 4000
VOLTS:
WATTS:
OUTPUT:
LINE FUSE:
Rosemount Analytical Inc.
Orrville, OH 44667-0901
85-264 VAC 48-62 Hz
TM
800-433-6076
4-20 mA
R
5 Amps
TM
HART
SMART FAMILY
CALIBRATION GAS
REFERENCE AIR
26170025
Figure 2-12. Oxymitter 4000 Gas Connections