![Rosemount 3144P User Manual Download Page 17](http://html1.mh-extra.com/html/rosemount/3144p/3144p_user-manual_1487451017.webp)
Reference Manual
00809-0100-4021, Rev FB
August 2011
2-3
Rosemount 3144P
Set the Switches
HART
Without an LCD display
1. If the transmitter is installed in a loop, set the loop to manual mode and
disconnect the power.
2. Remove the housing cover on the electronics side of the transmitter.
Do not remove the transmitter cover in explosive atmospheres with a
live circuit.
3. Set the switches to the desired position (see Figure 2-1).
4. Replace the transmitter cover. Both transmitter covers must be fully
engaged to meet explosion-proof requirements.
5. Apply power and set the loop to automatic mode.
With an LCD display
1. If the transmitter is installed in a loop, set the loop to manual mode and
disconnect the power.
2. Remove the housing cover on the electronics side of the transmitter.
Do not remove the transmitter cover in explosive atmospheres with a
live circuit.
3. Remove the housing cover, then unscrew the LCD display screws and
gently slide the meter straight off.
4. Set the switches to the desired position (see Figure 2-1).
5. Gently slide the LCD display back into place, taking extra precautions
with the 10 pin connection.
6. Replace and tighten the LCD display screws to secure the LCD display.
7. Replace the transmitter cover. Both transmitter covers must be fully
engaged to meet explosion-proof requirements.
8. Apply power and set the loop to automatic mode.
F
OUNDATION
Fieldbus
Without an LCD Display
1. If the transmitter is installed in a loop, set the loop to Out-of-Service
(OOS) mode (if applicable) and disconnect the power.
2. Remove the electronics housing cover.
3. Set the switches to the desired position and reattach the housing cover.
4. Apply power and set the loop to In-Service mode.
With an LCD Display
1. If the transmitter is installed in a loop, set the loop to Out-of-Service
(OOS) (if applicable) and disconnect the power.
2. Remove the electronics housing cover.
3. Unscrew the LCD display screws and pull the meter straight off.
4. Set the switches to the desired position.
5. Reattach the LCD display and electronics housing cover (consider LCD
display orientation).
6. Apply power and set the loop to In-Service mode.