Reference Manual
300550EN, Rev AA
December 2010
Rosemount 2240S
6-16
Section 6. Service and Troubleshooting
Device error
Input register no. 102.
Bit 0: RAM Error
Bit 1: FPROM Error
Bit 2: HREG Error
Bit 3: SW Error
Bit 4: Other memory error
Bit 6: Reserved for Display error
Bit 7: Reserved for Modem error
Bit 9: Internal temperature error
Bit 10: Other HW error
Bit 11 : Measurement Error
Bit 12: Configuration error
See “Device Errors” on page 6-18 for more
information on the various error types.
Contact Emerson Process
Management/Rosemount TankGauging
service department for support.
Device warning
Input register no. 104.
Bit 0: RAM Warning
Bit 1: FPROM Warning
Bit 2: HREG Warning
Bit 3: SW Warning
Bit 4: Other memory warning
Bit 6: Reserved for Display warning
Bit 7: Reserved for Modem warning
Bit 9: Internal Temperature Warning
Bit 10: Other HW warning
Bit 11: Measurement warning
Bit 12: Config warning
See “Device Warnings” on page 6-17 for more
information on the various warnings.
Contact Emerson Process
Management/Rosemount TankGauging
service department for support.
Internal temp
Input register no. 106. Internal
temperature.
Contact Emerson Process
Management/Rosemount TankGauging
service department for support.
Internal temp status
Input register no. 108.
Bit 0: Internal temperature out of limits
Bit 1: Temperature Device Error
Bit 15: Internal temperature not valid
MinIntTemp
Input register no. 110. Minimum
measured internal temperature.
MaxIntTemp
Input register no. 112. Maximum
measured internal temperature.
(1) The register number refers to the internal Input Register in the 2240S database.
Message
Description
Action
Summary of Contents for 2240S
Page 2: ......
Page 4: ......
Page 84: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S A 6 Appendix A Reference Data ...
Page 99: ......