Reference Manual
300550EN, Rev AA
December 2010
Rosemount 2240S
6-14
Section 6. Service and Troubleshooting
Configuration can not be saved
Write protection switch is set to the ON
position
• Check write protection switch on the 2240S, see “DIP
The Status LED is blinking error
codes
2240S transmitter, temperature element,
or Water Level Sensor error
• Check the 2240S for hardware or software errors.
• Check temperature elements.
• Check Water Level Sensor.
• See “Device Error LED Signals” on page 6-7.
• See “Device Errors” on page 6-18.
• Check Device Status input register (see “Device Status”
• Contact Emerson Process Management/Rosemount
TankGauging service department.
2240S icon in TankMaster
WinSetup is red
Simulation mode active
• Stop Simulation mode by setting the Simulate switch to
OFF(see “DIP Switches” on page 5-13).
• Stop Simulation mode in TankMaster WinSetup (open
WinSetup
Set Simulation Mode
window and click the
Stop button).
Symptom
Possible cause
Action
Summary of Contents for 2240S
Page 2: ......
Page 4: ......
Page 84: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S A 6 Appendix A Reference Data ...
Page 99: ......