2
Mount the transmitter
2.1 Liquid applications
Procedure
1. Place taps to the side of the line.
2. Mount beside or below the taps.
3. Mount the transmitter so the drain/vent valves are oriented upward.
2.2 Gas applications
Procedure
1. Place taps in the top or side of the line.
2. Mount beside or above the taps.
February 2019