MODEL 54 C
SECTION 3.0
PROGRAMING THE ANALYZER
SECTION 3.0
PROGRAMING THE ANALYZER
The analyzer arrives from the factory configured and ready to operate
for use with inductive (toroidal) con-
ductivity sensors.
If Series 100 contacting conductivity sensor is used the analyzer must be configured as
“contacting” under sensor type as in
Table 3-1. Conductivity Settings List
ITEM
RANGES
FACTORY SETTINGS
USER SETTINGS
PROGRAM LEVEL (Sections 3.1 - 3.2)
1. Alarm 1 (low action)
0 - 2,000 mS/cm
20 µS/cm
_______
2. Alarm 2 (high action)
0 - 2,000 mS/cm
1,000 mS/cm
_______
3. Alarm 3 (high action)
0 - 2,000 mS/cm
1,000 mS/cm
_______
1. Output 1: 4 mA (min value)
0 - 2,000 mS/cm
0 µS/cm
_______
2. Output 1: 20 mA (max value)
0 - 2,000 mS/cm
1,000 mS/cm
_______
3. Output 2: 4 mA (min value)
–15 - 200 °C
0.0 °C
_______
4. Output 2: 20 mA (max value)
–15 - 200 °C
100.0 °C
_______
CONFIGURE LEVEL (Sections 3.3 - 3.10)
A. Display (
1. Sensor type
Inductive/Contacting
Inductive
_______
2. Reference temperature
0 to 200°C
25°C
3. Temperature Units
°C/°F
°C
_______
4. Output 1 Units
mA/% of full scale
mA
_______
5. Output 2 Units
mA/% of full scale
mA
_______
6. Language
English/Français/Español/Deutsch/Italiano
English
_______
7. Operating display lower left
See section 3.3
Alarm 1 Setpoint
_______
8. Operating display lower right
See section 3.3
Alarm 2 Setpoint
_______
9. Display contrast
0-9 (9 darkest)
4
_______
B. Outputs (
1. Hold (Outputs and Relays)
Disable/Enable/ 20 min timeout
Disable feature
_______
2. Output 1
(a) Current Range
4-20 mA/0-20 mA
4-20 mA
_______
(b) Dampening
0-299 Sec
0 Sec
_______
(c) Hold Mode Value
Last value/Fixed value
Last value
_______
(value if (c) Fixed)
0-22 mA
21 mA
_______
(d) Fault value
0-22 mA
22 mA
_______
This section provides detailed procedures for changing
operational values for specific applications. Refer to
if a contacting conductivity sensor is used.
Before attempting any changes refer to the parameter
setup list shown in Table 3-1. This table presents a brief
description and the possible options.The factory set-
ting is also given with a space for the user setting. It is
recommended that the list be carefully reviewed before
any changes are made.
On initial configuration, it is recommended that the
parameters be entered in the order shown on the
worksheet. This will reduce the chance of accidentally
omitting a needed variable.