
Remote Commands
R&S
®
FPL1000
824
User Manual 1178.3370.02 ─ 09
Example:
INIT:CONT OFF
Stop the continuous sweep.
CALC:SGR:FRAM:SEL -25
Selects frame number -25.
Manual operation:
See
CALCulate<n>:SGRam:HDEPth
<History>
CALCulate<n>:SPECtrogram:HDEPth
<History>
This command defines the number of frames to be stored in the R&S
FPL1000 mem-
ory.
Suffix:
<n>
.
Parameters:
<History>
The maximum number of frames depends on the number of
sweep points.
Range:
781 to 20000
Increment: 1
*RST:
3000
Example:
//Set the history depth to 1500
CALC:SGR:SPEC 1500
Manual operation:
See
CALCulate<n>:SGRam:LAYout
<State>
CALCulate<n>:SPECtrogram:LAYout
<State>
This command selects the state and size of spectrograms.
The command is available for result displays that support spectrograms.
Suffix:
<n>
.
Parameters:
<State>
ON
Spectrogram and trace diagram share a window.
OFF
Only the trace diagram is displayed, the spectrogram is not.
*RST:
OFF
Example:
CALC4:SPEC:LAY FULL
Shows the spectrogram in window 4. The corresponding trace
diagram is hidden.
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters