Remote Commands
R&S
®
FPL1000
618
User Manual 1178.3370.02 ─ 11
<m>
<sb>
irrelevant
Parameters:
<Mode>
WRITe | MAXHold
WRITe
The power is calculated from the current trace.
MAXHold
The power is calculated from the current trace and compared
with the previous power value using a maximum algorithm.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
<Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
irrelevant
<m>
irrelevant
<sb>
Multi-SEM: 1 to 3
for all other measurements: irrelevant
Parameters:
<Measurement>
ACPower | MCACpower
ACLR measurements (also known as adjacent channel power or
multicarrier adjacent channel measurements).
Returns the power for every active transmission and adjacent
channel. The order is:
•
power of the transmission channels
•
power of adjacent channel (lower,upper)
•
power of alternate channels (lower,upper)
The unit of the return values depends on the scaling of the y-
axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
CN
Carrier-to-noise measurements.
Returns the C/N ratio in dB.
CN0
Carrier-to-noise measurements.
Returns the C/N ratio referenced to a 1
Hz bandwidth in
dBm/Hz.
Configuring and Performing Measurements