Remote Commands
R&S
®
FPL1000
794
User Manual 1178.3370.02 ─ 09
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
<RefValue>
This command defines the reference value for relative measurements.
Suffix:
<n>
.
<p>
Power sensor index
Parameters:
<RefValue>
Range:
-200 dBm to 200 dBm
*RST:
0
Default unit: DBM
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation:
See
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<n>
.
<p>
Power sensor index
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Usage:
Event
Manual operation:
See
"Setting the Reference Level from the Measurement Meas -
CALCulate<n>:PMETer<p>:RELative:STATe
<State>
This command turns relative power sensor measurements on and off.
Suffix:
<n>
.
<p>
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Setting Basic Spectrum RF Measurement Parameters