![Ronde & Schwarz 1304.0004.03 User Manual Download Page 753](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985753.webp)
Remote Commands
R&S
®
FPL1000
753
User Manual 1178.3370.02 ─ 09
<WindowType>
Type of result display you want to use in the existing window.
See
on page 750 for a list of availa-
ble window types.
Example:
LAY:REPL:WIND '1',MTAB
Replaces the result display in window 1 with a marker table.
Usage:
Setting only
LAYout:SPLitter
<Index1>, <Index2>, <Position>
This command changes the position of a splitter and thus controls the size of the win-
dows on each side of the splitter.
Note that windows must have a certain minimum size. If the position you define con-
flicts with the minimum size of any of the affected windows, the command will not work,
but does not return an error.
Figure 11-1: SmartGrid coordinates for remote control of the splitters
Setting parameters:
<Index1>
The index of one window the splitter controls.
<Index2>
The index of a window on the other side of the splitter.
Remote Commands for Result Displays