![Ronde & Schwarz 1304.0004.03 User Manual Download Page 666](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985666.webp)
Remote Commands
R&S
®
FPL1000
666
User Manual 1178.3370.02 ─ 09
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the SEM sweep list covers, the R&S
FPL1000 will not mea-
sure the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FPL1000 will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value. Note that the minimum frequency range of a
SEM range is 20
Hz.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a SEM range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the SEM sweep list covers, the R&S
FPL1000 will not mea-
sure the ranges that are outside the span - results may be invalid.
●
greater than the span the SEM sweep list covers, the R&S
FPL1000 will adjust the
start frequency of the first SEM range and the stop frequency of the last SEM
range to the span
For more information see
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Manual operation:
See
Configuring and Performing Measurements