![Ronde & Schwarz 1304.0004.03 User Manual Download Page 619](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985619.webp)
Remote Commands
R&S
®
FPL1000
619
User Manual 1178.3370.02 ─ 09
DISPlay:ATAB
<State>
This command switches between the MultiView tab and the most recently displayed
channel setup. If only one channel setup is active, this command has no effect.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
INSTrument:CREate:DUPLicate
This command duplicates the currently selected channel setup, i.e creates a new chan-
nel setup of the same type and with the identical measurement settings. The name of
the new channel setup is the same as the copied channel setup, extended by a con-
secutive number (e.g. "IQAnalyzer" -> "IQAnalyzer 2").
The channel setup to be duplicated must be selected first using the
INST:SEL
com-
mand.
Example:
INST:SEL 'IQAnalyzer'
INST:CRE:DUPL
Duplicates the channel setup named 'IQAnalyzer' and creates a
new channel setup named 'IQAnalyzer2'.
Usage:
Event
Manual operation:
See
INSTrument:CREate[:NEW]
<ChannelType>, <ChannelName>
This command adds an additional measurement channel. You can configure up to 10
measurement channels at the same time (depending on available memory).
See also
●
●
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
<ChannelName>
String containing the name of the channel.
Note that you can not assign an existing channel name to a new
channel; this will cause an error.
Example:
INST:CRE SAN, 'Spectrum 2'
Adds an additional spectrum display named "Spectrum 2".
Manual operation:
See
Selecting the Operating Mode and Application