![Ronde & Schwarz 1304.0004.03 User Manual Download Page 239](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985239.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
239
User Manual 1178.3370.02 ─ 09
Figure 7-36: Intermodulation products Ps1 and Ps2
The intermodulation product at f
i2
is generated by mixing the 2nd harmonic of useful
signal P
U2
and signal P
U1
.
The intermodulation product at f
i1
is generated by mixing the 2nd harmonic of useful
signal P
U1
and signal P
U2
.
f
i1
= 2 × f
u1
– f
u2
(1)
f
i2
= 2 × f
u2
– f
u1
(2)
Dependency on level of useful signals
The level of the intermodulation products depends on the level of the useful signals. If
the two useful signals are increased by 1 dB, the level of the intermodulation products
increases by 3 dB, which means that the spacing a
D3
between intermodulation signals
and useful signals is reduced by 2 dB. This is illustrated in
.
Measurements and Results