Continued…..
Check List
If your powerchair fails to operate, please check the following:
The unit is switched on.
All plugs and connections are firmly fixed.
Battery level shows full charge.
Freewheeling device is in the drive position
.
Circuit breaker has not tripped.
Battery and controller lead are properly connected.
If the freewheel device has been operated with the powerchair switched on, the
powerchair will not operate until the unit has been turned off and on again.
VR2 Error Codes
Your VR2 controller has the ability to diagnose possible errors in the electrical system,
which may prevent your powerchair from operating. These are displayed on the battery
gauge. The fault is shown against the number of lights flashing on the display.
Battery Gauge Possible Fault
10 bars flash
Battery voltage too high. Check battery connections.
9 bars flash
Electromagnetic brake fault – check freewheel.
8 bars flash
Possible controller fault – check connection / cable.
7 bars flash
Possible joystick fault – check connection / cable.
6 bars flash
Battery charger connected.
5 bars flash
Right hand motor wiring fault – check connection.
4 bars flash
Right hand motor disconnected – check connection.
3 bars flash
Left hand motor wiring fault – check connection.
2 bars flash
Left hand motor disconnected – check connection.
1 bar flashing
Low battery voltage – charge batteries.
Note: If your powerchair fails to operate after checking the above possible faults consult
your local Dealer immediately. Also ensure that the VR2 has not been accidentally
locked – refer to page 10.