Configuration
R&S
®
ZNL
40
User Manual 1178.5989.02 ─ 06
account manually, select "Manual" and enter the number in the "Number of Read-
ings" field.
8. To activate the duty cycle correction, select "DutyCycle" and enter a percentage as
the correction value.
9. If you selected "dB" or "%" as units (relative display), define a reference value:
a) To set the currently measured power as a reference value, press the "Meas ->
Ref" button.
b) Alternatively, enter a value manually in the "Reference Value" field.
c) Optionally, select the "Use Ref Level Offset" option to take the reference level
offset set for the analyzer into account for the measured power.
10. If necessary, repeat steps 3-10 for another power sensor.
11. Set the "Power Sensor State" at the top of the "Power Sensor" tab to "On" to acti-
vate power measurement for the selected power sensors.
The results of the power measurement are displayed in the marker table (Function:
"Sensor <1...4>" ).
How to Zero the Power Sensor
1. To display the "Power Sensor" tab of the "Input" dialog box, do one of the following:
● Select "Input" from the "Overview".
● Select the [INPUT/OUTPUT] key and then the "PowerSensorConfig" softkey.
2. Select the tab that is assigned to the power sensor you want to zero.
3. Press the "Zeroing Power Sensor" button.
A dialog box is displayed that prompts you to disconnect all signals from the input
of the power sensor.
4. Disconnect all signals sending input to the power sensor and press [ENTER] to
continue.
5. Wait until zeroing is complete.
A corresponding message is displayed.
5.3.4
Output Settings
Access
: "Overview" > "Output"
The R&S
ZNL can provide signals to different output connectors.
For details on connectors, refer to the R&S
ZNL Getting Started manual, "Front / Rear
Panel View" chapters.
Receiving Data Input and Providing Data Output