Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
181
User Manual 1178.5989.02 ─ 06
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM
Example:
CALC:THR -82DBM
Sets the threshold value to -82 dBm.
Manual operation:
See
CALCulate<n>:THReshold:STATe
<State>
This command turns a threshold for the marker peak search on and off (for
all
markers
in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:THR:STAT ON
Switches on the threshold line.
Manual operation:
See
" Deactivating All Search Limits "
9.7.3.4
Marker Search (Spectrograms)
The following commands automatically define the marker and delta marker position in
the spectrogram.
Using Markers
The following commands control spectrogram markers.
Useful commands for spectrogram markers described elsewhere
The following commands define the horizontal position of the markers.
●
CALCulate<n>:MARKer<m>:MAXimum:LEFT
●
CALCulate<n>:MARKer<m>:MAXimum:NEXT
●
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
●
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
●
CALCulate<n>:MARKer<m>:MINimum:LEFT
●
CALCulate<n>:MARKer<m>:MINimum:NEXT
●
CALCulate<n>:MARKer<m>:MINimum[:PEAK]
●
CALCulate<n>:MARKer<m>:MINimum:RIGHt
I/Q Analysis