Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
163
User Manual 1178.5989.02 ─ 06
CALCulate<n>:SPECtrogram[:STATe]
.............................................................................. 165
CALCulate<n>:SGRam:THReedim[:STATe]
...................................................................... 166
CALCulate<n>:SPECtrogram:THReedim[:STATe]
..............................................................166
....................................................................................... 166
CALCulate<n>:SPECtrogram:TRACe
...............................................................................166
CALCulate<n>:SGRam:TSTamp:DATA?
...........................................................................166
CALCulate<n>:SPECtrogram:TSTamp:DATA?
.................................................................. 166
CALCulate<n>:SGRam:TSTamp[:STATe]
..........................................................................167
CALCulate<n>:SPECtrogram:TSTamp[:STATe]
................................................................. 167
CALCulate<n>:SGRam:CLEar[:IMMediate]
CALCulate<n>:SPECtrogram:CLEar[:IMMediate]
This command resets the spectrogram and clears the history buffer.
Suffix:
<n>
.
1..n
irrelevant
Example:
CALC:SGR:CLE
Resets the result display and clears the memory.
Manual operation:
See
CALCulate<n>:SGRam:CONTinuous
<State>
CALCulate<n>:SPECtrogram:CONTinuous
<State>
This command determines whether the results of the last measurement are deleted
before starting a new measurement in single sweep mode.
This setting applies to all spectrograms in the channel setup.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
INIT:CONT OFF
Selects single sweep mode.
INIT;*WAI
Starts the sweep and waits for the end of the sweep.
CALC:SGR:CONT ON
Repeats the single sweep measurement without deleting the
results of the last measurement.
I/Q Analysis