Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
159
User Manual 1178.5989.02 ─ 06
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<Aperture>
Range:
1 to 50
*RST:
2
Default unit: PCT
Example:
DISP3:TRAC2:SMO:APER 5
Defines an aperture of 5% for trace 2 in window 3
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a particular trace on and off.
If enabled, the trace is smoothed by the value specified using
WINDow<n>][:SUBWindow<w>]:TRACe<t>:SMOothing:APERture
For more information see the R&S
ZNL User Manual.
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DISP3:TRAC2:SMO ON
Turns on trace smoothing for trace 2 in window 3
[SENSe:]AVERage<n>:TYPE
<Mode>
This command selects the trace averaging mode.
Suffix:
<n>
.
1..n
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
I/Q Analysis