Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
141
User Manual 1178.5989.02 ─ 06
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
" Changing the Automatic Measurement Time ( Meastime-
[SENSe:]ADJust:CONFigure:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
ZNL performs a measurement
on the current input data. This command selects the way the R&S
ZNL determines the
length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
ZNL determines the measurement length automatically
according to the current input data.
MANual
The R&S
ZNL uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation
*RST:
AUTO
Manual operation:
See
" Resetting the Automatic Measurement Time ( Measti-
See
" Changing the Automatic Measurement Time ( Meastime-
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
<Threshold>
When the reference level is adjusted automatically using the
on page 142 command, the internal attenuators are also adjusted. In order to
avoid frequent adaptation due to small changes in the input signal, you can define a
hysteresis. This setting defines a lower threshold the signal must fall below (compared
to the last measurement) before the reference level is adapted automatically.
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18
dBm.
Manual operation:
See
Configuring I/Q Analyzer Measurements