Administrative Tasks
R&S
®
ZND
64
Getting Started 1316.2407.02 ─ 52
7.2
Firmware Installation
Upgrade versions of the analyzer firmware are supplied as single executable setup
files (
*.exe
).
Administrator account
You need administrator rights to install a new firmware version. See note on
accounts and password protection"
To perform a firmware update:
1. Copy the setup file to any storage medium accessible from the analyzer. This can
be either the internal mass storage drive, an external storage medium (USB mem-
ory stick, external CD-ROM drive) or a network connection (LAN).
The default name of the internal drive is
C:
. External storage devices are automati-
cally mapped to the next free drive, i.e.
D:
,
E:
etc.
2. Run the setup file from the Windows
®
Explorer. Follow the instructions of the setup
wizard.
Setup files can be reinstalled.
Factory calibration
A firmware update does not affect the factory calibration.
Downgrade to a firmware version < 3.00
To downgrade the firmware from a version ≥ 3.00 to a version < 3.00, it is required to
uninstall the firmware using Windows 10 "Apps & features" (Windows 7 "Programs and
Features") before proceeding with the installation.
7.3
Remote Operation in a LAN
A LAN connection is used to integrate the analyzer into a home/company network.
LAN connectivity offers several applications, e.g.:
●
Transfer data between a controller and the analyzer, e.g. to run a remote control
program.
Remote Operation in a LAN