Command Reference
R&S
®
ZNB/ZNBT
980
User Manual 1173.9163.02 ─ 38
Setting parameters:
<Characterization>
Location of the created characterization / cal kit file (string
parameter, extension *.calkit):
If a path is specified, the file is saved to the analyzer's hard disk.
The default directory (
MMEMory:CDIRectory
) is
not
used.
If only the file name is specified, the file is saved to the calibra-
tion unit's internal (flash) memory. The factory calibration data
on the unit is not overwritten.
If the file name is prefixed by "SD:", the file is saved to the SD
card inserted at the calibration unit (if available/accessible).
Example:
CORR:COLL:AUTO:CKIT 'AutoCalChar'
Generate a cal kit file
AutoCalChar.calkit
for the active cali-
bration unit and store it in the calibration unit.
SENSe:CORRection:COLLect:AUTO:CKIT 'SD:test1'
Generate a cal kit file
test1.calkit
for the active calibration
unit and store it in the calibration unit.
Usage:
Setting only
Manual operation:
See
"File Name/ Comment (Optional)"
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
<Password>
Enters a password to enable a single password-protected action in the automatic cali-
bration or in the characterization wizard. If password protection has been activated
manually in the "Characterize Cal Unit" dialog, the password is required for any single
execution of one the following commands, provided that a cal kit (characterization) file
different from the active characterization is specified:
[SENSe]:CORRection:COLLect:AUTO
[SENSe]:CORRection:COLLect:AUTO:CKIT
[SENSe]:CORRection:COLLect:AUTO:PORTs
[SENSe]:CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe]:CORRection:COLLect:AUTO:TYPE
Tip:
You have to send
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
repeatedly if your command script uses several password-protected commands. You
do not need a password to perform automatic calibrations using the active cal unit
characterization.
Setting parameters:
<Password>
Password (string parameter), as defined in manual control.
SCPI Command Reference