Command Reference
R&S
®
ZNB/ZNBT
968
User Manual 1173.9163.02 ─ 38
<StandardType>
MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen |
FOPen | MSHort | FSHort | MOSHort | MOSHort1 | MOSHort2 |
MOSHort3 | FOSHort | FOSHort1 | FOSHort2 | FOSHort3 |
MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | FSMatch
Standard type; see
<DelayMode>
DELay | ELENgth
ELENgth (default) – ZVR compatible modelling
DELay – Agilent modelling
<Port1>[, <Port2>]
Optional port restriction: one port number for one port standards,
two port numbers for two port standards
Usage:
Query only
[SENSe:]CORRection:CKIT:<ConnType>:SELect
<CalKitName>
Selects the calibration kit to be used for a specified connector type
<ConnType>
. The
kit is identified by its name.
Tip:
For connector types with arbitrary, user-defined names you can use the command
[SENSe:]CORRection:CKIT:SELect
.
Cal kit naming conventions
Calibration kit names must be entered as string parameters. The string contains the cal
kit name used in the calibration dialogs (e.g. "Calibration Presettings"); a "Ω" in the
name must be replaced by
Ohm
, e.g.:
●
NewKit1
denotes the user-defined calibration kit "NewKit1".
●
N 50 Ohm Ideal Kit
denotes the "N 50 Ω Ideal Kit".
●
ZV-Z21 typical
denotes the cal kit "ZV-Z21 typical".
Suffix:
<ConnType>
.
Connector type, one of the following identifiers:
N50, N75
–
N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292
–
PC 7, PC 3.5 or 2.92 mm connectors
USER<no>
–
user-defined connectors UserConn1, UserConn2
SMA
–
user-defined connector type SMA ...
Parameters:
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer. See "Cal kit naming conventions " above.
*RST:
n/a (A *RST does not change the assignment
between connector types and calibration kits.)
SCPI Command Reference