Command Reference
R&S
®
ZNB/ZNBT
945
User Manual 1173.9163.02 ─ 38
Note that the decimal separator and field separator must be different: if both are set to
comma, actually a semicolon will be used as field separator.
Setting parameters:
<TraceName>
The name of the eye diagram whose results shall be exported
<TraceFile>
Mandatory string parameter containing the path and file name of
the file. If the path is omitted, the current directory is used (see
<DecSeparator>
POINt | COMMa
Decimal separator
*RST:
COMMa
<FieldSeparator>
SEMicolon | COMMa | TABulator | SPACe
Field separator
*RST:
SEMicolon
Usage:
Setting only
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
MMEMory:STORe:LIMit
<TraceName>, <LimLineFile>
Saves the limit lines associated to a specified trace to a limit line file. Limit lines are
created using the
CALCulate<Chn>:LIMit...
commands.
Setting parameters:
<TraceName>
Name of an existing trace in the active recall set (string parame-
ter) for which a limit line definition exists.
<LimLineFile>
String parameter to specify the name and directory of the cre-
ated limit line file. The default extension (manual control) for limit
line files is *.limit, although other extensions are allowed.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
Example:
See
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:MARKer
<AsciiFile>
Saves the values of all markers to an ASCII file.
SCPI Command Reference