Command Reference
R&S
®
ZNB/ZNBT
880
User Manual 1173.9163.02 ─ 38
Suffix:
<Wnd>
.
Number of the diagram area to become the active diagram area.
DISPlay:WINDow<Wnd>:MAXimize
acts on all diagrams of
the current recall set, however, the diagram no.
<Wnd>
is dis-
played on top of the others.
Parameters:
<Boolean>
ON | OFF - maximize all diagram areas or restore the previous
display configuration.
*RST:
OFF
Example:
*RST; :DISP:WIND2:STAT ON
Create diagram areas no. 1 (with default trace) and 2 (with no
trace).
DISP:WIND2:MAXimize ON
Maximize the diagram areas, placing area no. 2 on top.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:OVERview[:STATe]
<Boolean>
Enables the zoom function with an additional overview window for the diagram no.
<Wnd>
or removes the overview window from a diagram.
Suffix:
<Wnd>
.
Number of the zoomed diagram area
Parameters:
<Boolean>
ON
–
activate the zoom window with overview window
OFF
–
remove the overview window
*RST:
OFF
Example:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM[:
Manual operation:
See
DISPlay[:WINDow<Wnd>]:NAME
<Name>
Defines a name for diagram area
<Wnd>
. The name appears in the list of diagram
areas, to be queried by
DISPlay[:WINDow<Wnd>]:CATalog?
Suffix:
<Wnd>
.
Number of the diagram area.
Parameters:
<Name>
String variable for the name.
Example:
See
DISPlay[:WINDow<Wnd>]:CATalog?
Manual operation:
See
SCPI Command Reference