Command Reference
R&S
®
ZNB/ZNBT
834
User Manual 1173.9163.02 ─ 38
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>[:
STATe]
<Boolean>
Enables or disables the deembedding function for single ended ports. It is allowed to
change the circuit model and its parameters while deembedding is enabled.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<Boolean>
ON - deembedding active
OFF - deembedding inactive
*RST:
OFF
Example:
CALC:TRAN:VNET:SEND:DEEM:TND CSL
Select the Serial C, shunt L circuit model for deembedding.
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3; :
CALC:TRAN:VNET:SEND:DEEM ON
Increase the resistance R2 for the Serial C, shunt L circuit model
to 2.2 kΩ and enable deembedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>
<CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>?
<CircuitModel>
Specifies the capacitance value C<Cmp> in the different circuit models for single
ended port embedding.
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of capacitance in circuit model. The total number of
capacitances depends on the selected circuit model.
Parameters:
<Capacitance>
Capacitance C<Cmp> for the specified circuit model.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F