Command Reference
R&S
®
ZNB/ZNBT
778
User Manual 1173.9163.02 ─ 38
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus
range between 2 GHz and 3 GHz, assigning a ripple limit of +5
dB.
CALC:RIPP:SEGM:COUNT?
Query the number of ranges. The response is
2
.
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
<Boolean>
Enables or disables the limit check in the ripple limit range no.
<Seg>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Number of the ripple limit range.
Parameters:
<Boolean>
ON | OFF - Limit check on or off. A result is available even if the
limit check is disabled; see example for
*RST:
n/a (no ripple limit line defined after a reset)
Example:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
Manual operation:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
<Limit>
Defines the ripple limit for ripple limit range no.
<Seg>
. A range must be created first to
"Rules for Ripple Test Definition"
Tip:
To define several ripple limit ranges with a single command, use
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Number of the ripple limit range.
SCPI Command Reference