Command Reference
R&S
®
ZNB/ZNBT
775
User Manual 1173.9163.02 ─ 38
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<RippleLimRange>
Parameter list in the format <Type>, <StartStimulus>, <StopSti-
mulus>, <RippleLimit>{, <Type>, <StartStimulus>, <StopStimu-
lus>, <RippleLimit>}, where:
<Type>
–
Boolean identifier for the ripple limit range type. 1 for
ripple limit range on (with limit check). 0 for ripple limit range off:
The range is defined, but no limit check result displayed. The
result is still available via
<StartStimulus>
–
stimulus value of the first point of the ripple
limit range. The unit is adjusted to the sweep type of the active
channel (
<StartStimulus>
–
stimulus value of the last point of the ripple
limit range. The unit is adjusted to the sweep type.
<RippleLimit>
–
ripple limit in the range. The unit is adjusted to
the format of the active trace (
).
Range:
Virtually no restriction for ripple limit ranges.
*RST:
n/a (no ripple limit line defined after a reset)
Default unit: Stimulus values: Hz for frequency sweeps, dBm for
power sweeps, s for time and CW mode sweeps.
Ripple limit: see above.
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus
range between 2 GHz and 3 GHz, assigning a ripple limit of +5
dB.
CALC:RIPP:DISP ON
Show the ripple limits in the active diagram.
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DELete:ALL
Deletes all ripple limit ranges.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference