GUI Reference
R&S
®
ZNB/ZNBT
630
User Manual 1173.9163.02 ─ 38
Redefining physical ports causes a factory reset and deletes all switch matrix RF con-
nections.
a wave, b wave, Source
Define a physical port by assigning its reference receiver, measurement receiver and
generator, respectively.
Remote command:
[SENSe:]UDSParams<Pt>:PARam
[SENSe:]UDSParams:ACTive
5.19.5.3
External Matrices Dialog
The "External Matrices" dialog allows to discover and register external switch matrices.
Access:
SYSTEM – SETUP > "External Ports" > "Switch Matrix..."
Background information
Chapter 4.7.20, "External Switch Matrices"
The registration of a new external switch matrix typically involves the following steps:
1. connect the switch matrix to your R&S
ZNB/ZNBT using LAN or USB interface
2. If the switch matrix is connected via LAN, enable "LAN detection"
3. tap "Scan Instruments" and wait until the switch matrix appears in the table of
"Known Devices"
4. tap
to copy the switch matrix to the list of configured devices
Setup Softtool