GUI Reference
R&S
®
ZNB/ZNBT
528
User Manual 1173.9163.02 ─ 38
Note:
Display of the limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:DISPlay[:STATe]
Limit Check
Switches the limit check of the active trace on or off.
When the limit check is switched on, a movable "PASS" or "FAIL" message is dis-
played in the diagram. If the limit check fails at a measurement point, the point is
marked with a colored square. An acoustic signal (
) and a TTL signal
indicating pass or fail can be generated in addition.
The appearance of the limit fail symbols is defined in the "Define User Color Scheme"
dialog (see
Chapter 5.17.3.2, "Define User Color Scheme Dialog"
can choose between various options:
●
Change the trace color between failed measurement points.
●
Show or hide the colored squares.
Note:
●
Limit check and display of the limit lines are independent of each other: With dis-
abled limit check, the limit line can still be displayed. If no limit lines are defined for
the active trace, the limit check can be switched on but will always "PASS" the
trace.
●
If enabled,
Figure 5-47: Simultaneous "dB Mag" limit line and circle check
Lines Softtool